Issue StoriesStreck Lab's Sugar-Chex Linearity tames matrix effect in POC glucose test stripsTaming the matrix effect in POC glucose strip testing By Coleen Curran For most of the population, the words matrix effect conjures up images of Keanu Reeves flying though steel walls in a floor-length leather coat, but for laboratorians it brings to mind something far less entertaining bad results. NCCLS defines matrix effect
Although matrix effect can occur across the spectrum of clinical laboratory tests, it tends to become a pronounced problem in hospitals that use a number of different types and brands of point-of-care glucose meters. To control each individual glucose meter correctly, the user commonly employs a brand-specific control. When one considers the number of glucose meters, test strips and control strips residing in the average hospital, its easy to see how things can quickly get out of hand. Verifying the linearity of glucose meters is a fact of life for point-of-care departments. CLIA mandates it, and the CAP survey requires that linearity checks be performed on each new meter before it is used on a patient and then twice a year on working glucometers. In determining a departments capability of capturing a true glucose, it is necessary to verify the linearity of the entire glucose monitoring system operator technique, reagent strip lot variation and instrument performance. Unlike glucose controls for wet chemistry analyzers, the suspending medium in controls for point-of-care glucose meters plays a major role in determining the final glucose concentration. Thus, the preparation of glucose controls for POC devices requires that the characteristics of whole blood be present in order to obtain accurate values. Early glucose controls Building a blood-like linearity check For example, glucose in H2O on YSI = 60 mg/dL. On a test strip, the same sample measures 175 mg/dL. Similar discrepancies are seen with serum. A Sugar-Chex II/Sugar-Chex Linearity glucose control gives good agreement with YSI and other instruments. (Yellow Springs Inc., is the Gold Standard glucose reference analyzer.) Glucose agreement between POC and the central lab Streck Labs has identified these key elements as: a cellular element, viscosity and ionic strength. The cellular element in whole blood contributes viscosity, which in turn affects the regulation of sample flow through the various strip layers. For example, red blood cells physically fill pores on the reagent membrane, thus regulating flow to the subsequent layer. Ionic strength impacts reaction kinetics that in turn shortens or extends the reaction phase. In the absence of either the cellular element or ionic strength, sample migration through the layers is markedly different. Rate of absorption, sample delivery through the reagent membrane and rate/speed of the enzymatic reaction is changed. Ionic strength, combined with close simulation of the cellular elements functional characteristics (size, density, charge and concentration) permit the control solution to function universally and maintain good agreement with the central laboratory. Bradford A. Hunsley, research and development manager at Streck Labs describes matrix effect as inaccurate measurement of the specified parameter. If you have a true glucose concentration of 100 mg/dL; a matrix effect would be the recovery of an excessive amount of or an insufficient amount of glucose. The true glucose concentration is being masked in one way or another, he said. Precision cannot replace accuracy Through research, Streck has identified the essential elements in blood that are relative to point-of-care glucose meter performance and constructed a matrix that closely mimics whole blood. To achieve this mimicry across multiple strip and meter platforms, it recognized the essential features associated with flow rate/viscosity, wetting properties and ionic strength. In combination, these elements produce a glucose composition that closely replicates the behavioral characteristics of fresh whole blood on point-of-care glucose monitoring technologies. Many commercially available controls do not contain the cellular element, Hunsley said. Given that these strips are designed to handle, manage and separate cellular materials, you can see that the flow characteristics and the reaction kinetics are altered when you apply a control material that the strip is not designed for. Baylor Medicals experience With the Level 5, youre trying to get a higher value and [Sugar-Chex Linearity] was able to produce a higher value, said Claudia Hill, point-of-care coordinator for Baylor University Medical Center at Waxahachie. The other product we were using didnt quite get it as high as we needed it to. Kathy Belote, point-of care coordinator at Baylor University Medical Center, Garland reported the same problem, We had trouble with Level 5 dropping off, she said. It was coming out lower than what we like. It was in the range, but when you graphed it, it looked like it was way off. So we and some of the other Baylor affiliate hospitals switched to Strecks product because we were all having the same problems. Like many busy point-of-care departments that strive for optimal patient care, Baylor realized that its not enough just to have a reproducible number. Reproducibility and precision must be coupled with accuracy. With Sugar-Chex Linearity, its point-of-care departments are confident that the glucose numbers recovered are meaningful, even in the very high and low ranges. Coleen Curran is a freelance writer and healthcare consultant based in Charlotte, N.C. |
|
|
Featured Jobs
Find a Job |
ADDITIONAL ONLINE RESOURCES |
Featured Employer
|