Issue StoriesMeasurement of XL-FDPS as a Diagnostic Aid in Thrombotic DisordersBy Robert Janetschek, MSc, MT (ASCP) & RS Greenfield, PhD
Biochemistry & Clinical Significance of XL-FDPs As the fibrin clot is formed, hemostatic mechanisms also activate the fibrinolytic system, which is responsible for clot lysis. Tissue plasminogen activator (tPA) cleaves plasminogen to form plasmin, the major clot-lysing protease that rapidly cleaves the cross-linked fibrin causing dissolution of the clot. Small peptidyl by-products, referred to as FDPs, are then produced by the proteolytic cleavage of the cross-linked fibrin clot by plasmin. There are two types of FDPs, non-cross-linked FDPs and cross-linked FDPs (XL-FDPs), where XL-FDPs contain fibrin fragments cross-linked together by the action of FXIIIa. Therefore, the presence of XL-FDPs in the blood is a reliable indicator of recent fibrin clot dissolution activity. XL-FDPs are increased in disseminated intravascular coagulation (DIC), in surgical and trauma patients, sickle-cell disease, liver disease, severe infection, sepsis, inflammation, malignancy, and pre-eclampsia coagulopathy. Elevated levels of XL-FDPs are found during myocardial infarction, and in some cases used as a predictive indicator. D-dimer, one of the more distinct and specific monomers of XL-FDPs, is a widely used selective marker for venous thromboembolisms (VTE). Levels of the XL-FDPs proteins may also increase during normal pregnancy, but very high levels are often associated with complications. Finally, quantitation of XL-FDPs has been used to monitor thrombolytic therapy with tissue-type plasminogen activator (tPA) and streptokinase (SK). However, the concentration of XL-FDPs detected in a specimen will depend on several factors, such as the severity of the thrombotic episode, the rate of cross-linked fibrin formation and degradation, and the time elapsed between the thrombotic event and drawing of the blood from the patient. Measurement of XL-FDPs However, advances in assay technologies have lead the way toward improvements in measuring XL-FDPs for use in the clinical laboratory. Sensitive, rapid, and convenient laboratory tests for XL-FDPs are now available. These tests greatly improve the diagnosis and patient management during a thrombotic crisis. For example, the development of highly specific monoclonal antibodies has lead to the development of qualitative, semiquantitative, and quantitative clinical diagnostic assays for these specific degradation products. As a result of these advances, American Diagnostica Inc of Stamford, Conn, recently introduced the new ACTISCREEN XL-FDP assay. ACTISCREEN XL-FDP is intended for use as a rapid qualitative, or semiquantitative, procedure to determine the presence of XL-FDPs in human plasma. The assay is a manual immunoagglutination test procedure utilizing latex beads coupled with a murine monoclonal antibody against XL-FDPs. The monoclonal antibody utilized in the assay is highly specific for XL-FDPs and does not cross-react with fibrinogen, factor XIIIa cross-linked fibrin, or non-cross-linked FDPs. In the simple, 3-minute test procedure, plasma is mixed with the antibody-coated latex beads on a conveniently designed solid support. The XL-FDPs present in the plasma sample will bind to the coated latex beads, resulting in the aggregation of the beads that is unmistakably visible to the human eye. Positive test results can be seen at concentrations of XL-FDPs in plasma above 200 ng/mL (0.20 mg/L), the lower limit of detection for the assay. This results in a highly sensitive test procedure. In addition, a semi-quantitative test result, if desired, can be obtained by testing serially diluted patient samples. Performance Characteristics ACTISCREEN XL-FDP and the titers based on ELISA obtained XL-FDP values were r = 0.91 for citrated samples, r = 0.73 for EDTA samples and r = 0.78 for heparin samples, respectively. Citrate is the obvious anticoagulant of choice. In specificity studies, plasma from 162 out of 170 apparently healthy, volunteer blood donors tested negative (95.3% specificity) using ACTISCREEN XL-FDP. With regard to precision, the intra-assay (within-run) reproducibility was determined for 10 replicates of three plasma samples that contained different levels of XL-FDPs. The results were found to be essentially equivalent for all replicates tested at each concentration of XL-FDP. Inter-assay (run-to-run) reproducibility was determined using ten plasma samples with XL-FDP titers ranging from 1 to 16. In 10 runs, the replicates of these specimens did not vary by more than one titer. Interference studies showed that substances such as bilirubin to 0.2 mg/mL, hemoglobin to 5.0 mg/mL, lipemia (triglycerides) to 30 mg/mL and protein (gamma globulin) to 0.06 g/mL do not affect the results with ACTISCREEN XL-FDP assay. Additional studies have suggested that ACTISCREEN XL-FDP is also insensitive to rheumatoid factor (RF) disturbances. Key Features & Benefits Additionally, the ACTISCREEN XL-FDP assay is sufficiently versatile to be used in a variety of clinical testing facilities and settings. For example, in POLs, large group practices, and freestanding clinics, ACTISCREEN XL-FDP could be employed as a testing device where the diagnostic profile calls for immediate on-site testing. Second, ACTISCREEN XL-FDP is also ideally suited for use in small hospitals, and STAT and satellite laboratories that do not have fully automated testing capabilities or that have limited access to sophisticated diagnostic equipment. The same holds true for regional core laboratories responsible for coordinating a group of smaller laboratories with limited instrumentation. Lastly, in commercial reference and large hospital laboratories, ACTISCREEN XL-FDP can be utilized as a backup to automated procedures, or as an after-hours test. Robert Janetschek, MSc, MT (ASCP) and RS Greenfield, PhD, can be reached c/o American Diagnostica Inc, PO Box 110215, Stamford, Conn 06911-0215; (203) 602-7777, Fax: (203) 602-2221. |
|
|
Featured Jobs
Find a Job |
ADDITIONAL ONLINE RESOURCES |
Featured Employer
|